27
S
7.6. Replacing Spring Arm for Wall Mounted Single Light
WARNING:
Before performing the following steps, the electricity for the Lights and/or
Monitors has to be powered OFF.
CAUTION:
There is a risk of injury from spring-loaded joints. If the spring arm is
released under tension, it will move quickly and uncontrollably, possibly causing
injury or damage to equipment.
1. Hold onto the spring arm when installing or removing the light head.
2. Avoid putting pressure on the spring arm and releasing it suddenly.
3. Spring arm should always have the stop set for the spring arm to be parallel.
1. Remove the Light Head as instructed in
section 7.9. Remove the Brake Screw (3).
2. Remove Contact Holder (1) by removing the
two Philips screws (2).
WARNING:
The spring arm WILL fall once the two Circlips are Removed.
3. Remove the two Circlips using Circlip Pliers.
4. Remove Shim rings.
5. Remove the Spring Arm.
6. Push New Spring Arm onto the Horizontal
Arm.
7. Install Brake Screw.
8. Install the two Shim Rings removed above in
Step 5.