120
S
Fuse block replacement:
• Turn off the main power at the circuit breaker
• Work with the hospital/electrician to ensure power to the room and the Chromophare system is
turned off. Carefully disconnect the Chromophare system from hospital mains.
WARNING: Failure to turn off the
main power to the Chromophare
system prior to disconnecting wires
can result in electrical shock.
If the hospital has the
new
fuse block configuration
(fused line with unfused neutral), the line fuse alone can
be replaced.
• To replace the fuse, turn the fuse cap
counterclockwise to remove.
• Install new fuse with same amp rating into cap
and turn clockwise to lock in place.
If the hospital has the
old
fuse block configuration
(fused line and fused neutral), follow these steps to
replace:
• Note where Chromophare wires are attached into
the blocks. Disconnect all internal Chromophare
wires from the set of blocks requiring
replacement.
For SK box systems:
• Remove the current clamps from the mounting
rail for the affected light by pinching the front clip
for easy release.
• For SK electronics, the line and neutral
must both be replaced for all lights on the
suspension.