122
S
• Using service kit P41494:
• Connect the 8 amp fuse (500393) to
the fuse clamp (39965) and secure by
twisting clockwise. Add end bar (39967)
to the fuse clamp.
• Mount the new Line, Neutral, and Ground
blocks to the DIN rail as described. The
arrows on the fuse clamps must be facing
the tube when electronics are tube
mounted.
For single light systems
Note:
Additional suspension items such as
flat panels may result in additional ground
blocks on the flange. Mimic the current
suspension setup when replacing
• Place the blocks in the order: Ground,
Neutral, Line, Ground as shown.
For dual light systems
• Place the blocks in the order: Ground, Neutral, Neutral, Line, Line, Ground as shown.
• Install a dual shorting bar across the two neutral blocks and secure with 2 screws. Install a dual
shorting bar across the two line blocks and secure with 2 screws.
For triple light systems:
• Place the blocks in the order: Ground, Neutral, Neutral, Neutral, Line, Line, Line, Ground as shown.
• Install a yellow colored triple shorting bar across the three neutral blocks and secure with 3 screws.
Install a brass colored dual shorting bar across the three line blocks and secure with 3 screws.