Instruction Manual for Sicuro pesa Hospital Bed
Page
79
9.4
A
MBIENT
C
ONDITIONS
The following ambient conditions must be maintained:
In storage/ transport:
Minimum
Maximum
Temperature :
-10° C
+ 50° C
Relative humidity :
20 %
90 %
At 30°C; not condensed
Air pressure:
800 hPa
1060 hPa At altitude of ≤ 2000m
In operation:
Minimum
Maximum
Temperature:
+ 5° C
+ 40° C
Relative humidity :
20 %
90 %
At 30°C; not condensed
Air pressure:
800 hPa
1060 hPa At altitude of ≤ 2000m
9.5
T
ECHNICAL
I
NFORMATION ON
E
LECTROMAGNETIC
C
OMPATIBILITY
(EMC)
To ensure EMC, only use cables and accessories approved by the manufacturer
(see Chapter 9.3).
The use of accessories, sensors or cables other than those approved, with the
exception of sensors and cables sold by the equipment manufacturer as
replacement parts for internal components, can result in an increase in the
transmission level or a reduction in the immunity level of the equipment.
The equipment may not be used directly next to or on top of other equipment.
If it is necessary to use the equipment in this way, you must check to ensure
that it functions properly in the required configuration.
Guidelines and Manufacturer’s Declaration – Electromagnetic Emissions –
The BED is intended for use in the electromagnetic environment described below. The operator
or user of the BED should ensure that it is used in such an environment.
Interference emission
measurements
Compliance
Electromagnetic environment guidelines
HF emissions
to CISPR 11
Group 1
The BED uses HF energy for its internal functions
only.
HF emissions
to CISPR 11
Class B
The BED is intended for use in all types of
establishment including residential and the like
that are directly connected to a public supply
network that also serves buildings that are used
for residential purposes.
Harmonics according to
IEC 61000-3-2
Class D
Voltage fluctuations/flicker
acc. to IEC 61000-3-3
Complies
HF emissions according to
CISPR 14-1
Complies
The BED is not intended for connection to other
technical equipment.