Instruction Manual for Sicuro pesa Hospital Bed
Page
71
PROBLEM
Possible Causes
Solution
Mains control
lamp in control
unit does not light
up
No power supply to socket
Mains cable damaged
Fuses in control unit
defective
Use a socket that works
properly
Replace the mains cable
Replace the control unit.
Inform the operator in order
to arrange for repairs
Drive runs for a
brief time only,
then stops
Drive overloaded
Remove the overload, retest
Control unit is not
functioning;
locking LEDs on
control box
alternate
between orange
and green
One or more motors are
not connected
There is a serious problem
with the control unit.
For safety reasons, all
functions are locked.
Connect all motors; reset
the control unit (RESET); see
Chapter 6.5.2.2
If problem recurs: have drive
system checked.
Inform the operator in order
to arrange for repairs
Height
adjustments and
tilting not possible
or only in one
direction;
acoustic signal
sounds during
adjustment
Control unit has “lost” the
drive positions
Initialise drives; see Chapter
6.5.2.3
CPR signal switch
not functioning:
When any button
is pressed on the
control box at the
foot end or the
patient handset,
three beeps
sound in addition
Faulty CPR switch;
Faulty MJB;
Plug-in contacts in the
cable connection are
faulty
Replace the CPR switch
Replace the MJB
Check plug-in
connections/replace cable