Instruction Manual for Sicurao pesa Hospital Bed
Page
56
6.1.2
Servicing Plan
Measure
Material/Tool
Check that all screw connections and/or metal
safety caps are firmly positioned;
tighten/replace if necessary.
Suitable tools.
If necessary, clean and slightly oil the
connecting bolts on the headboard and
footboard as well as the corresponding sleeves
on the vertical bed frame tubing.
Resin-free and acid-free
grease.
If necessary, clean and lightly spray the pivot
and friction bearings.
Resin-free and acid-free oil
spray (e.g. lubricant A).
Check the adjustment and routing of the
Bowden cable for the emergency release of
the backrest. No sharp bends or kinks.
Contact Stiegelmeyer service
personnel.
Rechargeable batteries must be charged and
fully functional.
Connect bed to mains supply.
Recommendation: We
recommend that
rechargeable batteries be
replaced after 5 years (wear
and tear part).
If the plugs have been disconnected (e.g.
when drives have been replaced or during
troubleshooting), always check to see if the O-
rings (sealing rings) of the electrical plugs are
present and/or damaged. The plugs of new
electrical components are always fitted with a
new O-ring.
Only replace O-rings with
original LINAK O-
rings.- (obtainable from
Stiegelmeyer Service; see
Chapter 6.4)
Check the corresponding
recesses in the plug for dirt or
damage before reinserting the
plugs.
Check earth wires for tears and tightness of
screws.
Replace torn or damaged
earth wires straps.
Touch up any damage in the coating.
Special-purpose paint
(mechanically stable
anticorrosive protection) in a
suitable colour; see following
page for details.
In order to optimise the fault-free operation of the beds and thus minimise
downtime, we recommend that servicing of the bed should include additional
special preventative measures.
A practical guide on preventative replacement of components that may be
affected can be obtained from our service centre.