Instruction Manual for Sicuro pesa Hospital Bed
Page
67
Drive for lateral tilting of mattress base
(version EM5 only)
Set the mid position
(only necessary in addition if the drive does not stop
automatically at the horizontal mattress base position)
Settings can be carried out any time during normal operation.
It is not necessary to carry out a RESET.
M4: Drive for lateral tilting of mattress base
(optional feature)
3x
+
Re-adjust the required horizontal mid
position (=reference position)
Press the memory button
briefly three times
and then
only
the button shown
A beep is emitted: Drive M4 is re-adjusted
to the mid position
6.5.2.4
Restore Factory Settings
Must be carried out
After a RESET
To return to the defined identical positions of the memory buttons.
Effect:
Resets the internal settings (e.g. position memory) of the CB16 to the
defined factory settings
Where?
On the control box [
13
] at the foot end.
Check for...
Press both buttons at the same time and keep them pressed until the
acoustic signal stops.