Instruction Manual for Sicurao pesa Hospital Bed
Page
70
7
Troubleshooting
7.1
T
ROUBLESHOOTING
G
UIDE
The following table is a guide for rectifying faults:
PROBLEM
Possible Causes
Solution
Handset/
Drive unit system
not functioning
Mains cable not plugged
in
No power supply to socket
Plug not inserted properly
Drives locked
Handset, mains cable or
control unit is defective
Insert mains cable; green
LED must light up
Check mains socket and/or
fuse box
Check plug connections
Enable functions
Inform the operator in order
to arrange for repairs
Handset / foot
pedal not
functioning,
adjustments are
enabled
Faulty handset
Foot pedal has not been
separately enabled
Replace handset.
Enable the foot pedal
(Chapter 0)
Inform your operator to
arrange for the necessary
repairs
Operation using
battery not
possible
Battery not charged
There is no rechargeable
battery
Connect the bed to the
mains supply for approx. 8
hours
Constant alarm
sounds during
adjustment
Battery capacity depleted
Connect bed to the mains
supply to recharge battery
as soon as possible
Operation with
sufficiently
charged battery
only possible for a
short time
Limit of battery’s service life
reached
Replace battery.
Inform the operator in order
to arrange for repairs
Operation is not
possible despite
proper power
supply
Control unit has shut off
due to over-heating
Control unit defective
Observe max. duty
cycle: ON/OFF 2/18 min
Replace the control unit.
Inform the operator in order
to arrange for repairs
Manual
emergency
lowering of
backrest is not
possible
Bowden cable is too loose
or not secured
Bowden cable is kinked
Readjust at the release lever
or secure
Install new Bowden
cable. Inform your operator
about any necessary repairs