Instruction Manual for Sicurao pesa Hospital Bed
Page
64
damage to the junction box electronics!
6.5.2.2
Resetting Control Unit CB16 to Original Settings (RESET)
Must be carried out
One or more locking LEDs flashing on control
box
After replacing CB16
------------------None-----------------
To reset after system errors, e.g.
faulty handset / internal fault in
CB16
All 5 LEDs flashing
If the adjustment drives lose their
position(if motorised adjustments
are no longer possible and only
intermittent beeping is heard)
One or more LED pairs M1; M3; M4 are
flashing
(see also Chapter 7.2, Error codes)
Effect:
- Deletes any errors in CB16 (reset)
- Deletes any lifting limits programmed
Where?
On the control box [
13
] at the foot end.
Version EM4
Version EM5
1.
Press and hold both
"RESET"
buttons
at the same time
2.
An
acoustic signal will sound.
3.
Release the buttons as soon as the signal stops after 5 seconds.
The positions of the memory buttons (sitting position, normal position, sleeping
position) are retained.
If motorised adjustments are still not possible after this,
and only intermittent beeping can be heard when the
adjustment buttons are pressed, and one or more of
the locking LEDs on the control box continues to flash:
Initialising Drives
(see following page for details)
M1 M3
M4
1
2
M1 M3 M4
----------ERROR-----------
-