Instruction Manual for Sicurao pesa Hospital Bed
Page
74
9
Technical Data
9.1
D
IMENSIONS AND
W
EIGHTS
Dimensions
Weights
External dimensions: approx. 216 x 1025 cm
Empty weight:
(without
accessories)
approx. 195kg
Mattress base:
(mattress
dimensions)
200 x 90 cm
Safe working
load:
260kg
Mattress base
sections:
80/20/40/60 cm
(back/seat/thigh/lower
leg)
Max. patient
weight:
220 – 250kg
(depending on the
weight of accessories
attached)
Castors:
Ø 15 cm
Protection height of
safety sides:
428 mm
(above the mattress base)
9.2
A
DJUSTMENT
O
PTIONS
Height adjustment
range:
Approx. 45 to 80 cm
Thigh rest:
to approx. 80°
Time for total
height adjustment
approx. 16 s
Tilting range of
mattress base to
Trendelenburg position:
to approx. 19°
Backrest:
Length
compensation
acc. to DbfK
to approx. 70°
approx. 80 mm
Tilting range of
mattress base to
Reverse-Trendelenburg
to approx. 20°
Lateral tilting of
mattress base (optional
feature)
to approx. +/-
25°
9.3
E
LECTRICAL
D
ATA
Mattress Base:
The mattress base is always equipped with a connection for potential
equalisation at the head end of the bed. The earth wire of the mains cable is
conductively connected to the mattress base frame through the control unit.
This qualifies the bed as a protection class I product.
Note:
If proper operation of the earth wire cannot be guaranteed, the bed may
only be operated with integrated batteries (if available), or it has to be
connected to the hospital's potential equalisation.