4-3
Cleaning The Equipment
Operator Manual
10043124
4.3 Areas To Be Cleaned
Before Each Use
NOTE: Refer to the user manual supplied with the monitor for the
manufacturer's recommended procedures for cleaning the monitor
and the monitor control surface.
The following areas of the system must be cleaned and disinfected
before each use of the lighthead (see
for instructions):
•
Suspension Arm –
Wipe the entire suspension arm, including
the suspension fork and yoke.
•
Lighthead –
Wipe both top and side surfaces.
•
Sterile Handle Support –
Wipe all areas of the support,
including those covered when the handle is installed.
•
Lens –
Important:
Clean only the exterior surface of the lens.
a.
Remove the lighthandle.
b. Clean/disinfect the outer surface of the lens as outlined in
c.
Wipe the outer surface of the lens with an anti-static cleaner
and soft cloth.
d. Do not reinstall a lighthandle until immediately before the
light is to be used in a surgical procedure.
•
When using a sterilizable lighthandle, verify that it is
sterilized according to facility sterile protocols between
each surgical procedure.
•
When using disposable sterile lighthandle covers, verify
a new one is used for each procedure.
NOTE: Always sterilize handle using conventional
hospital sterilization procedures and a standard
sterilization prevacuum or gravity cycle.
•
Wall Control –
Wipe panel with an anti-static cleaner and soft
cloth.
WARNING–PERSONAL INJURY
and/or POSSIBLE EQUIPMENT
DAMAGE HAZARD: Do not attempt
to clean lighthead unless power is
turned off and the lighthead has
cooled sufficiently.
CAUTION – POSSIBLE EQUIPMENT
DAMAGE HAZARD:
•
Cleaning and disinfecting
agents used on this lighting
system must be certified by their
manufacturer to be compatible
with the following materials:
polycarbonate, polyetherimide,
santoprene.
•
DO NOT SPRAY any cleaning
product directly onto the
lighthead or any system
components.
•
Use only recommended
cleaning/disinfecting and/or
anti-static agents on this light.
Staining, pitting and/or
discoloration could occur if a
phenolic-, iodophor-, or
glutaraldehyde-based
disinfectant is used on the
surfaces of this light. Also, use
of alcohol or aerosol spray
cleaner/disinfectants containing
a substantial amount of alcohol
in the formula can damage the
polycarbonate lens.
•
Do not scratch the lens when
cleaning; always wear rubber
gloves and use only a clean,
white, lint-free cloth when
wiping external surfaces.
•
Prevent leakage of fluids into
interior of lighthead or wall
control. Any such leakage could
impair or damage the lighting
system.
•
Do not use floor cleaners on this
equipment.
!
!
Summary of Contents for HarmonyAIR G Series
Page 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Page 2: ......
Page 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Page 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Page 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Page 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Page 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...