1-5
Safety Precautions
Operator Manual
10043124
Avoid discoloration of control center keypad and display. Do not clean control center with povidone-iodine
solutions or allow such solutions to contact keypad and display surfaces.
Do not bump lightheads into walls or other equipment.
To avoid inadequate balancing, load added to single monitor yoke must not exceed 15 kg (33 lb).
To avoid inadequate balancing, load added to dual monitor yoke must not exceed 10.5 kg (23 lb) per
monitor.
To avoid excessive radiated emissions on camera equipped systems, utilize a maximum 3’ (1 m) shielded
HDMI cable with one Fair-Rite shield bead (0446176451) and a single turn of cable at the HDMI output
and a Fair-Rite shield bead (0444164181) clamped to the distal end of the HDMI cable. Shield bead
characteristics for 0446176451 are 85 ohms at 10 MHz, 137 ohms at 25 MHz, 330 ohms at 100 MHz and
360 ohms at 250 MHz. Characteristics for 044176451 are 138 ohms at 25 MHz and 230 ohms at 100
MHz. Alternately, use a HDMI cable that has been verified to meet CISPR 11 Class A when connected to
the scaler output.
CAUTION – POSSIBLE EQUIPMENT DAMAGE: (CONTINUED)
!
!
!
!
!
Summary of Contents for HarmonyAIR G Series
Page 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Page 2: ......
Page 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Page 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Page 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Page 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Page 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...