Operator Manual
Operating Instructions
3.
Sterilize handle and silicone ring separately (disassembled, but
in the same chamber) using a cycle with either of the following
parameters:
Prevacuum
Temperature:
270°F (132°C)
Exposure Time:
4 minutes
Gravity
Temperature:
270°F (132°C)
Exposure Time:
10 minutes
NOTE: Before using silicone ring, note the following:
1. Reuse limit is 1,000 cycles.
2. Replace ring if yellowing, deformation, cracking or other signs of
deterioration are present.