2-5
Installation Verification
Operator Manual
10043124
2.5 Environmental
Conditions
2.6 Power
Requirements
Three Lighthead System (system running)
For sealed-ceiling applications
Control Center Ambient
= 60°C (140°F) Maximum
Room Ambient
= 20 to 25°C (68 to 77°F)
Interstitial space temperature has minimal effect on the canopy ambient in a completely sealed ceiling
application.
For vented-ceiling applications (1/4" [6 mm]
minimum gap between ceiling and ceiling plate)
Control Center Ambient
= 60°C (140°F) Maximum
Room Ambient
= 20 to 25°C (68 to 77°F)
To avoid adverse thermal effects on canopy control electronics, thermal contributions from interstitial space
and room ambient temperatures must
NOT
result in canopy ambient temperatures exceeding 60°C (140°F).
Recommended Transport/Storage Temperature
(not exceeding 15 weeks)
2 to 38°C (36 to 100°F)
Recommended Relative Humidity
30 to 75%
Atmospheric Pressure
500 to 1060 hPascals
RPC: Interstitial space temperature when RPC is
mounted above ceiling
45°C (113°F)
Two-Lighted System
100 – 240 Vac, 50/60 Hz
5-2A
500 Watts at 100 Vac
480 Watts at 240 Vac
Typical Lighthead Wattage
40 Watts
Summary of Contents for HarmonyAIR G Series
Page 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Page 2: ......
Page 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Page 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Page 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Page 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Page 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...