2-3
Installation Verification
Operator Manual
10043124
2.3 HarmonyAIR
Surgical Lighting
System G5 Lighthead
Optical Performance
Important:
Values are typical for the small pattern size at highest
intensity setting (unless otherwise noted) at 100 cm (39-3/8") from
the lighthead. Definitions and measurements are in accordance with
IEC 60601-2-41.
Table 2-1. HarmonyAIR Surgical Lighting System G Series Lighthead Optical Performance
Feature
G5 Lighthead
Maximum
Central Illuminance
160,000 lux
Peak Total
Irradiance
< 500 W/m
2
Pattern Size
Fixed
18 cm (7")
Adjustable
18 cm (7") – 28 cm (11")
D50 Diameter
62% of small beam pattern
(see Pattern Size)
Depth of Illumination (to 20%)
(to 60%)
97 cm (38")
53 cm (21")
Color Temperature (CCT)
4,400K ± 300K
General Color Rendering Index (CRI)
Up to 96
Deep Saturated Red Color Rendering
Index (R9)
Up to 98
Shadow Control
Single Mask
44%
Double Mask
43%
Cavity
100%
Single mask w/ Cavity
44%
Double mask w/ Cavity
43%
LED Life
50,000 hours at full intensity
Summary of Contents for HarmonyAIR G Series
Page 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Page 2: ......
Page 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Page 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Page 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Page 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Page 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...