![Status Pro BG 840300 User Manual Download Page 38](http://html1.mh-extra.com/html/status-pro/bg-840300/bg-840300_user-manual_1367494038.webp)
38
Status
Pro
–
µLine F1 User Manual
PERFORMING MEASUREMENT
8 Performing Measurement
of Accuracy in Position
8.1 Reading Data
Press 'Start' in the positioning menu to active measurement. The tooling machine will
move along the points of measurement according to its CNC programing. If this is done
manually, the operator must make sure to pass the starting and ending points as
described above (see fig. 41). Also, at each measuring point allow enough time for the
point to be captured by the system to guarantee a accurate measurement. You can see
the current status in the software and check with the control unit on the machine.
When a measuring point is successfully logged in the system a message will appear (see
fig. 44). The order of phase is as follows:
Positioning
➜
Axis within window frame (according to parame-
ters set in 7.2.3 and number of points in the list)
➜
Software pend-
ing (usually caused by vibration along the axis)
➜
Value being
logged in (wait for message 'saved' in green)
➜
Proceed to next
step.
Fig. 44: Notice alert
after point is success-
fully 'logged in'