Status
Pro
–
µLine F1 User Manual
21
FEATURES OF THE SYSTEM BUILD
Using the tripod, you can align the interferometer furthermore. Fig. 12 shows the different
handles and knobs for fine and coarse adjustment.
Here too, be sure to center
the fine adjusting before
alignment. Fine adjustment
follows rough adjustment.
Procedure for vertical alignment
Set approx. height by adjusting legs
➜
Use the handle on the tripod's mounting adapter
(see fig. 12) and fasten with the lock.
➜
Use the fine adjust on the mounting adapter for the
interferometer (fig. 12).
Procedure for horizontal alignment
Move tripod horizontally to the reference axis
➜
Use the fine adjust on the interferometer's
mounting adapter (fig. 12).
Procedure for tilt alignment
Manually arrange the angle of the pivot (tilt coarse pivot in fig. 12) on the mounting and
fasten
➜
Fine tune with screws and knob on the interferometer's slide if need be (fig. 11).
Using the tripod and adapters makes alignment easier. Without it, possibilities remain limited
and virtually no fine adjustment can be performed.
Fig. 12: Adjustments on the Tripod
Handle for
height adjust
Lock
Pivot for
coarse tilt
Horizontal
fine adjust
Vertical
fine adjust