General Information
xi
Temperature sensors
Excitation range
10, 100, 1000
μ
A
Thermistors
0.010
500
k
Ω
RTD
Pt-100, Pt-1000, and equiv
IC voltage sensors (1 mA bias)
LM335 and equiv.
IC current sensors (
>
5 V bias)
AD590 and equiv.
Measurement & display
Temperature resolution
0.001
◦
C
Thermistor resolution
0.1
Ω
Accuracy [4]
±
0
.
2 %
+
0.05
Ω
TEC current resolution
1
Accuracy
±
5
mA
TEC voltage resolution
1
Accuracy
±
5
mV
Conditions:
[1] Accuracy figures represent the uncertainty added by the measure-
ment circuitry inside the instrument. Sensor uncertainty is not in-
cluded, which can add up to 2
◦
C.
[2] Measured by a monitor thermistor on the TEC stage. The LDC500 is
operated in an environmental chamber between 0
◦
C and 40
◦
C. The
TEC and heatsink are at room temperature.
[3] Measured output current noise from 10 Hz to 1 MHz.
[4] With 1 mA excitation current
General characteristics
Interface
GPIB (IEEE-488.2), RS-232, Ethernet
/
TCP-IP
Connectors
DB9-F (laser diode), DB15-F (TEC),
BNC (modulate in, trigger out)
Power
100 W, 100 V, 120 V, 220 V, 240 V
50 Hz or 60 Hz
Fuse
Type 5MF,
5
×
20 mm, “fast blow”
4 A (100 V
−
120 V); 2 A (220 V
−
240 V)
Weight
15 lbs
Dimensions
8
.
5
W
×
5
H
×
15
D
LDC500 Series Laser Diode Controllers