4 – 44
Remote Operation
LM335 Slope Parameter
TLMS(?)
{
f
}
Set (query) the slope parameter for LM335 calibration
{
to
f
}
, in
◦
C
/
V.
The LM335 calibration model follows
T
=
s
·
V
+
c
, where
V
is the
raw sensor voltage.
TLMS
sets the
s
coe
ffi
cient in the calibration.
Front panel:
#
d
>
.
TLMS?
Example:
1.02
LM335 O
ff
set Parameter
TLMY(?)
{
f
}
Set (query) the o
ff
set parameter for LM335 calibration
{
to
f
}
, in
◦
C.
The LM335 calibration model follows
T
=
s
·
V
+
c
, where
V
is the
raw sensor voltage.
TLMY
sets the
c
coe
ffi
cient in the calibration.
Front panel:
@0)
d
.
TLMY?
Example:
-2.02
AD590 Slope Parameter
TADS(?)
{
f
}
Set (query) the slope parameter for AD590 calibration
{
to
f
}
, in
◦
C
/μ
A.
The AD590 calibration model follows
T
=
s
·
I
+
c
, where
I
is the raw
sensor current.
TADS
sets the
s
coe
ffi
cient in the calibration.
Front panel:
#
d
>W*
.
TADS?
Example:
1.0
AD590 O
ff
set Parameter
TADY(?)
{
f
}
Set (query) the o
ff
set parameter for AD590 calibration
{
to
f
}
, in
◦
C.
The AD590 calibration model follows
T
=
s
·
I
+
c
, where
I
is the raw
sensor current.
TADY
sets the
c
coe
ffi
cient in the calibration.
Front panel:
@0)
d
.
TADY -273.15
Example:
LDC500 Series Laser Diode Controllers