4 – 2
Remote Operation
4.1
Index of Common Commands
Symbol
Definition
f
,
g
,
h
Floating-point value
i
,
j
Unsigned integer
z
Literal token
(?)
Required for queries; illegal for set commands
var
Parameter always required
{
var
}
Required parameter for set commands; illegal for queries
[
var
]
Optional parameter for both set and query forms
Laser limits
SVLM(?)
{
f
}
4 – 18 Laser Diode Voltage Limit
SILM(?)
{
f
}
4 – 18 Laser Diode Current Limit
PILM(?)
{
f
}
4 – 19 Photodiode Current Limit
PWLM(?)
{
f
}
4 – 19 Photodiode Power Limit
Laser settings
LDON(?)
{
z
}
4 – 20 Laser Diode Current On
/
O
ff
SILD(?)
{
f
}
4 – 20 Laser Diode Current Setpoint
SIPD(?)
{
f
}
4 – 20 Photodiode Current Setpoint
SWPD(?)
{
f
}
4 – 21 Photodiode Power Setpoint
SCAN(?)
{
f, i, g
}
4 – 21 Multi-step Scan of Laser Setpoint
SYND(?)
{
h
}
4 – 22 Sync delay for scans
Laser monitor
RILD?
4 – 22 Laser Diode Current Reading
RVLD?
4 – 23 Laser Diode Voltage Reading
RIPD?
4 – 23 Photodiode Current Reading
RWPD?
4 – 23 Photodiode Power Reading
ILOC?
4 – 24 Interlock
Laser configuration
RNGE(?)
{
z
}
4 – 24 Laser Diode Current Range
MODU(?)
{
z
}
4 – 24 Modulate Enable
SMOD(?)
{
z
}
4 – 25 Laser Control Mode
SIBW(?)
{
z
}
4 – 25 Laser Modulation Bandwidth
PDMW(?)
{
z
}
4 – 26 Photodiode Milliwatt Units
BIAS(?)
{
f
}
4 – 26 Photodiode Bias Voltage
CALP
f
4 – 27 Calibrate Photodiode Responsivity
RESP(?)
{
f
}
4 – 27 Photodiode Responsivity
SMLK(?)
{
z
}
4 – 28 Lock Laser Control Mode When On
LDC500 Series Laser Diode Controllers