Si->Swir1 Transition Wvl
When in “Remove Overlap” mode, Sets
the point where the Si detector data ends,
and the first InGaAs detector data begins,
when the “Remove Overlapped Detector
Data” option is selected.
Swir1->Swir2 Transition Wvl
When in “Remove Overlap” mode, Sets
the point where the first InGaAs detector
data ends, and the second InGaAs
detector data begins, when the “Remove
Overlapped Detector Data” option is
selected.
Matching Type Frame
Controls the method of matching Si and
InGaAs detector data.
None
No matching take place. When this
option is in effect, none of the other
Matching parameters has any effect on
spectral data.
Radiance
Uses the detectors’ radiance curve data to
perform matching on both the Reference
and Target scans separately.
Reflectance
Combine the Reference and Target scans
into a reflectance curve, and uses this
curve’s data to perform matching.
Matching Region Frame
Defines the region over which the Si and
first InGaAs data is examined in order to
perform detector matching.
Beginning Wvl
The beginning wavelength in the
matching region. This settings has no
effect if “Matching Type” is set to
“None”
Ending Wvl
The ending wavelength in the matching
region. This setting has no effect if
“Matching Type” is set to “None”.
Matching Factor Limits Frame
Defines the range of acceptable values
for the factor used to match Silicon data
to SWIR data.
Min
The minimum matching factor value.
Max
The maximum matching factor value.
Use NIR-SWIR Overlap Algorithm Frame
Use The NIR-SWIR Overlap Algorithm…
When checked, the Silicon and SWIR1
data are combined (within the Overlap
Region) using a combination of the two