Manual V2.docx | Status 10.12.2015 | Version 1.0
Figure 44. Screenshot of the Device Controls window for the Spin Detector Control in SpecsLab Prodigy.
Table 12. Typical parameter values for the Ferrum Detector when operated in spin mode.
Parameter
Value (V)
Scattering Energy
6.5
Lens 1
220
Lens 2
21.5
Lens 3
415
Lens 4
49.5
Lens1DeflectorX
0
Lens1DeflectorY
0
Lens1Stigmator
0
Lens3DeflectorX
0
Lens3DeflectorX
0
Lens3Stigmator
0
Due to the geometry of the Ferrum detector, it is to be expected that the optimum values are
not the same as for operation with the total intensity channel.
A.4 Alignment with phosphor screens
The following procedures are normally only carried out at SPECS. They are described here
for reference purposes.