![SPECS 2D-CCD User Manual Download Page 36](http://html1.mh-extra.com/html/specs/2d-ccd/2d-ccd_user-manual_1343051036.webp)
35
Manual V2.docx | Status 10.12.2015 | Version 1
Figure 16. The intensity for a kinetic energy of 40 eV as a function of the pass energy, E
p
.
VLEED spin detector
5.5
For information about the Ferrum spin detector and its principle of operation, please refer to
the Focus Ferrum Spin Detector manual.
PCU 300 Electronics
5.6
The PCU 300 is mounted directly onto the CEM flange of the Ferrum detector, in order to
receive the signals from the CEMs without any cable losses. It serves as a counter and
preamplifier.
The PCU 300 is a SPECS design. The Focus Ferrum Spin Detector manual describes a different
design. You should bear this in mind when reading the Focus Ferrum spin detector manual.
Example of experimental results
5.7
Graphene on Ir(111) has Rashba split surface states with opposite spin.
2
In addition, the
system is stable enough to be exposed to air. This makes it suitable as a test system for spin
detectors. In the experiments described below, ARUPS with HeII excitation (excitation energy
2
A. Varykhalov, D. Marchenko, M. R. Scholz, E. D. L. Rienks, T. K. Kim, G. Bihlmayer, J. Sánchez-Barriga, and O. Rader, PRL 108,
066804 (2012).
Summary of Contents for 2D-CCD
Page 1: ...User Manual Status 10 12 2015 Version 1 0 User Manual V1 0 October 19 2015...
Page 42: ...41 Manual V2 docx Status 10 12 2015 Version 1 Figure 22 The component spectra...
Page 53: ...52 Manual V2 docx Status 10 12 2015 Version 1 Figure 26 Quantum well state of 12 ML Pb Cu 111...
Page 57: ...56 Manual V2 docx Status 10 12 2015 Version 1 Figure 29 Wiring diagram for hardware type 156...
Page 85: ...84 Manual V2 docx Status 10 12 2015 Version 1 0...