37
Manual V2.docx | Status 10.12.2015 | Version 1
Figure 18. The surface states probed using the HAD mode. The spin detector integrates the signal over the angle
range indicated in the figure. Lens mode LAD, pass energy 20 eV.
The spin detector integrates the signal over the angle range indicated in Figure 18. By
using a circular hole as entrance aperture, or by using the iris aperture of the
PHOIBOS analyzer, it is possible to reduce the angle acceptance range to a
minimum of 0.3°. With an entrance aperture 3 mm in diameter, an angle
acceptance range of 1° is obtained, as shown in Figure 19.