![SPECS 2D-CCD User Manual Download Page 14](http://html1.mh-extra.com/html/specs/2d-ccd/2d-ccd_user-manual_1343051014.webp)
13
Manual V2.docx | Status 10.12.2015 | Version 1
9.
Set the voltages according to the values in Table 1 (the first line has already been
completed in the previous steps).
Table 1. Ramping up detector values
Ramp time
(s)
Udet actual
(V)
Udet target
(V)
Uscrn actual
(V)
Uscrn target
(V)
600
0
0
0
1000
1800
0
1000
1000
0
4000
1000
1000
0
2000
4000
1000
1500
2000
2500
2.4
Conditioning the spin detector
Following the bakeout, you need to prepare a fresh iron film on the tungsten substrate in the
detector. This procedure is described in the Focus Ferrum Spin Detector manual — please see
"Preparing of the Fe film on the W(100) crystal" in the Operation chapter.