5.
Successively reduce the entrance aperture size and iris opening while adjusting the
electron gun deflectors so that the intensity is optimal.
6.
At the end optimize the signal with the Slit6/open set and the iris fully closed.
The detector image will be a narrow stripe at angle zero, as shown in Figure 40. Check that
the stripe gets wider in symmetrical fashion, when the iris is opened.
Figure 40. Detector image with slit6/open and iris fully closed.
This means that the beam of the electron gun hits the sample on the electron optical axis of
the PHOIBOS lens.
A.3.2 Adjusting the deflector/rotator lens
When you have aligned the electron gun correctly, you can adjust the deflectors in the
deflector/rotator lens system. Measure the intensity with the total intensity channel of the
Ferrum detector as described below:
1.
Connect to the PHOIBOS 1D analyzer and the SpinDetectorControl in Device
Controls in Prodigy.
2.
Set the electron gun at the optimum position found in Section A.3.1.
3.
In the Experiment Editor, open the Template PHOIBOS 1D Adjustment.
4.
Start a fixed energy scan for E
kin
= 40 eV and E
p
= 20 eV. Operate the analyzer in the
LAD mode. Enable the overwrite mode.
Summary of Contents for 2D-CCD
Page 1: ...User Manual Status 10 12 2015 Version 1 0 User Manual V1 0 October 19 2015...
Page 42: ...41 Manual V2 docx Status 10 12 2015 Version 1 Figure 22 The component spectra...
Page 53: ...52 Manual V2 docx Status 10 12 2015 Version 1 Figure 26 Quantum well state of 12 ML Pb Cu 111...
Page 57: ...56 Manual V2 docx Status 10 12 2015 Version 1 Figure 29 Wiring diagram for hardware type 156...
Page 85: ...84 Manual V2 docx Status 10 12 2015 Version 1 0...