3-9
FVS-1000 MMP1
3-3-3. Cleaning the FLC Assembly
1.
Remove the gate base (35) assembly.
(Refer to section 3-2-1.)
2.
Remove the integrating sphere from the gate base (35)
assembly.
3.
Loosen the four screws, and pull out the lamp house
assembly. (Refer to section 3-4-1.)
4.
Loosen the three screws (with stopper), and remove
the lamp house cover. (Refer to section 3-4-1.)
Integrating sphere
(1) Blow off the dusts on the glass surface of the integrat-
ing sphere (2 locations) using the air blower for optical
use.
(2) For stubborn dusts which cannot be removed using the
air blower, clean with a cleaning cloth moistened with
cleaning liquid for optical use.
n
After cleaning, install the integrating sphere after the
cleaning liquid has evaporated completely.
F-FLC
n
Do not push the glass side of the F-FLC with force or hit it
with a hard object as this may cause the performance of the
FLC assembly to drop considerably, or disable use.
(1) Blow off the dusts on the glass side of the F-FLC
using the air blower for optical use.
(2) For stubborn dusts which cannot be removed using the
air blower, clean with a cleaning cloth moistened with
cleaning liquid for optical use.
n
After cleaning, install the lamp house assembly after
the cleaning liquid has evaporated completely.
Integrater assembly
Gate base (35) assembly
Glass
Glass
F-FLC
Lamp house assembly
3-3. Cleaning the Optical System