PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
33
4
Measurements
NOTE!
The socket marked with
symbol is connected with the PE pin of the test socket. It
is not allowed to apply dangerous voltage to this socket.
Tested appliance must be turned on.
Continuous measurement lasts until
STOP
is pressed.
After completing each measurement its parameters, date and duration of the meas-
urement can be viewed.
4.1
Visual check (preliminary test)
Select
Visual check
.
Examine the power cord and
the mains plug for cracks and
overheated spots.
The icon
will take you to
the menu where you can add
and remove photos from the
inspection. Adding photos is
done using the
Sonel PAT
Analysis Mobile
application
and the phone
’s camera.
If everything is in order:
select
All OK
or
tick
Visual test is posi-
tive
and go further using
icon
.