PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
49
4.7
Measurement of touch leakage current I
T
Press
I
T
.
Before starting the measurement, similarly to
sec. 4.2
, following parameters must be set:
test dura-
tion
,
limit
,
polarity
and
method
.
In
Change polarity
select:
yes
if the measurement is
to be repeated for reverse
polarity,
no
if the measurement is
performed for only one po-
larity.
In
Method
select:
Probe
if the measurement
will be carried out using
probe, and go to step
,
PRCD
if you want to test a
device with built-in RCD,
and go to step
Measurement with probe
Connect the mains plug of the tested appliance into the test socket of the tester. Use the
probe connected to
T2
terminal socket and touch the accessible parts of the tested appli-
ance (for Class I appliances - touch accessible parts not connected to PE).