PAT-80
● PAT-85 ● PAT-86 – USER MANUAL
58
Measurement without clamp
Connect the mains plug of the tested appliance into the test socket of the tester.
Press
START
.
Upon completing the measurement read the result. You can finish the test
before the defined test time duration by pressing
STOP
.
Compare the result with the technical data of the tested appliance. The assessment of the correct-
ness of the test results can be performed by selecting the proper field in ‘
Positive test result
’ or
‘
Negative test result
’. When saving test results in the memory, this assessment will also be saved
along with the results.
NOTE!
During the measurement, the same mains voltage is present at the measuring socket
which powers the tested appliance.
If the tested appliance is damaged, signaling a 16 A fuse burnout may also mean that
the overcurrent protection device in the mains from which the meter is powered has
tripped.