Chapter 4
Installation and Start up
2. Select the icon
Another window appears as shown below.
3. Select the icon
and the software will end.
4. Turn off the monitor.
4.4.3 Shutting down for 3 or more days inactivity
Upon completion of the working day whereas the analyzer will be
inactive for a period of more than two days (consisting also of
prolonged weekends) the following steps should be completed:
It is possible to carry out this procedure for short periods of inactivity
too(for example, during weekends).
1. When the work session is finished:
a.
Rinse and fill two empty starter reagent bottles with distilled
H
2
O, preferably at 35° - 40°C.
b.
Replace the starter reagents bottles with the two bottles of
distilled H
2
O.
c.
Replace the Wash/System Liquid tank with a tank of distilled
H
2
O.
d.
Execute “
System Test
” selecting the values at least equal to
the following (see Fig
4.4.3-1
):
Fig 4.4.3-1: Decontamination Priming Table
Section
Type
Values
Cycles
Pipettor
3
Washer
3
Chamber Set A
3
Reaction modules
BGW
0
LC - le
0
LC - ri
0
Note: The System Test function of <LC-le> and <LC-ri> is compressed into the <LC> for the MAGLUMI
1000
2. Follow the normal shutting down procedure (see section 4.4.2).
Operating instructions
Page 4-11