Chapter 4
Installation and Start up
4.3 Starting the MAGLUMI® software & system
The normal icon for entering the MAGLUMI® software is located
directly on the windows desktop.
When entering the software anytime the analyzer has been shut off,
it must be ensured that the analyzer remains off until prompted to
do so.
1. Double click on the icon in figure 4.4-1 by using either the mouse
or the touch screen.
Fig 4.3-1: MAGLUMI® software access example
2. Before the software can be accessed, a user name and
password must be given. The user can change the software
access to accept any name and password wished (see
Chapter 6 section 6.4).
Fig. 4.3-2 User login
3. Enter the user name & password. The software will start.
4. When the message “
No connection to device
” appears, turn
on the analyzer and wait until the LED’s of all free tracks in the
patient area have changed from red to green.
5.
Afterwards
press
<OK>
. The analyzer will begin to clear all
working channels in the system to ensure no collisions occur
during normal operation.
The system is ready when no further pop-ups appear on the monitor.
Operating instructions