Safety Notes
5 Danger of Fire
5.1 Handling with Flammable Chemicals
When using flammable chemicals such as organic solvents, there is
the possibility of catching fire.
For keeping safety, only use wash buffers, starter reagents
and diagnostic kits approved by SNIBE to assure not using
flammable chemicals in the working process.
This instrument is not explosion-proof type. Do not use organic
solvent in the direct surrounding of the analyzer whose ignition point
is lower than 65°C.
6 Danger of Explosion of Flammable Vapor
6.1 Handling with Flammable Chemicals
When flammable chemicals such as organic solvents are used in the
laboratory, there is the possibility of rising vapor concentrations
exceeding the explosion limit concentration, explosion can be
caused.
When using chemicals in the laboratory such as organic solvent,
which are flammable and easy vaporized, take care of leakages and
puddles, and ventilate the room sufficiently.
For keeping safety, only use wash buffers, starter reagents
and diagnostic kits approved by SNIBE to assure not using
flammable chemicals in the working process.
This instrument is not explosion-proof type. Do not use organic
solvent in the direct surrounding of the analyzer whose ignition point
is lower than 65°C.
7 Danger of Electric Shock
7.1 Electric Shock caused by touching the Analyzer Inside
When removing Cover of the MAGLUMI
®
Analyzer to work for
replacing parts, cleaning etc., turn off the power, disconnect the
Power Cord without fail.
Fig. 7.1-1: Safety Label nearby Power Connection
Operating instructions
Page 5