Chapter 3
System description
Fig. 3.2.2-2: Reagent station with 15 tracks for Reagent Integrals MAGLUMI 1000/2000.
Operating the reagent station:
Open the swivel flap of the reagent station only briefly for loading
purposes to maintain the refrigerated ambient of the reagent station.
Remove the protection foil from the Reagent Integral. When the RFID
Tag is put near the sensing area of the RFID reader ,a ”tick” soundsas
it has been read correctly
Take hold of the Reagent Integral (prepared according to the
manufacturer’s instructions) by its handle and slide it into the selected
reagent station track. Insert the integral up to the limit stop. A correctly
inserted integral is recognized by the software and displayed on the
monitor screen.
Always allow the Reagent Integral to remain in the reagent station for
a minimum of 30 minutes before use. During this time, the magnetic
particles in the analyzer will be automatically set in motion and
completely re-suspended.
Before using the Reagent Integrals, read the IFU (instructions
for use) provided in the reagent package (storage, preparation)!
Operating instructions
Page 3-9