Chapter 4
Installation and Start up
4.1.2 Connecting supply containers for wash / system liquid and waste containers
The supply connections for wash / system liquid (MAGLUMI
®
Wash / System Liquid) are on the right side of the analyzer. The
system liquid should be prepared in accordance with the delivered
IFU (
I
nstructions
F
or
U
se).
Container Connections
1. Install the System liquid cap to the container marked “
System
Liquid
”
2. Install the Waste cap to the liquid container marked “
Waste
”
3. Connect the two tubes of the system liquid cap to the analyzer
connections marked “
System Liquid 1
”. & “
System Liquid 2
”.
4. Connect the two tubes of the waste liquid cap to the analyzer
connections marked “
Waste Liquid 1
” & “
Waste Liquid 2
”.
5. Insert the cable connectors of the system liquid cap into the
receptacle marked “
System Liquid Sensor
”.
6. Insert the cable connector of the waste liquid cap into the
receptacle marked “
System Liquid Sensor
”.
4.1.3 Connecting starter reagents (MAGLUMI
®
Starter Kit)
The reagent reservoir contains the starter reagents (MAGLUMI
®
Starter Kit) and is located on the right side of the analyzer.
Starter Reagent Connections
1. Remove the protection cover from the starter reagent reservoir.
2. Place the screw connection with the white tube labeled “S 1“, on
the bottle marked “
MAGLUMI
®
Starter 1
”.
3. Place the screw connection with the white tube labeled “S 2“,, on
the bottle marked “
MAGLUMI
®
Starter 2
”.
4. Connect the sensor cable labeled “S 1“ of the starter reagent
reservoir to the pin on the cap of the bottle marked “
MAGLUMI
®
Starter 1
”.
5. Connect the sensor cable labeled “S 2“ of the starter reagent
reservoir to the pin on the cap of the bottle marked “
MAGLUMI
®
Starter 2
”.
6. Reinstall the cover for the starter reagent reservoir.
Page 4-4
Operating instructions