Safety Notes
Fig. 8.4-1: Warning Labels on Waste Bag housing
9 Affected processing of MAGLUMI
®
Diagnostic System
1.For keeping safety and correctness of the
MAGLUMI
®
Diagnostic System the Daily Maintenance and
System Test with Visual inspection of needle is to be
performed by the user every morning and after each
exchange of starter reagents.
Additionally the Routine weekly maintenance must be
performed by user according to the Operating Instructions.
2.For keeping safety and correctness of the diagnostic results,
the laboratory has to use MAGLUMI
®
Controls according to
their Information of Use.
3.The Information of Use of the used Diagnostic Kits are to be
followed. The MAGLUMI
®
Diagnostic System may only be
used under the working conditions defined in this Operating
Instruction.
The detailed cleaning instruction in the Operating Instructions
are to be followed strictly inclusively the choice of disinfectant.
4.Safe and intended function of the MAGLUMI
®
Diagnostic
System can only be expected with the use of cuvettes,
integrals, disposals and other acessories approved by SNIBE
5.Pooled Starter Systems can cause higher uncertainties in the
creation of the diagnostic results; therefore, Starter Reagents
may never be pooled.
6.Kits with different Starter Reagents may never be mixed on
board of the MAGLUMI
®
Analyzer.
Operating instructions
Page 9