Safety Notes
1 Caution on Electromagnetic Wave Interference
1.1 Electromagnetic Wave Interference given by MAGLUMI
®
to other Equipment
The use of the MAGLUMI
®
Analyzer may interfere with radio
and television reception.
Use the cables attached at the installation for connection between
the devices in the system. The proper use of the specified cables
minimizes Electromagnetic Wave Interference.
Installation and service of the system or changes in the
installation may never be performed by persons not authorized by
SNIBE, especially never connect devices in the system with
movable multiple plug sockets.
However, there is no guarantee that the MAGLUMI
®
Analyzer will
not cause Electromagnetic Wave Interference.
a. When the MAGLUMI
®
Analyzer may be the cause, turn off the
power of this Instrument and check radio and television reception.
b. If it is improved, the MAGLUMI
®
Analyzer probably is the cause.
1.2 Electromagnetic Wave Interference that may be given to the MAGLUMI
®
Analyzer
If the MAGLUMI
®
Analyzer is used near equipment that generates
strong electric and magnetic field, noises may enter the Instrument
to affect the performances and functions.
Use the cables attached at the installation for connection between
the devices in the system. The proper use of the specified cables
minimizes electromagnetic wave interference.
Installation and service of the system or changes in the
installation may never be performed by persons not authorized by
SNIBE, especially never connect devices in the system with
movable multiple plug sockets.
However, there is no guarantee that the MAGLUMI
®
Analyzer will
not be affected by Electromagnetic Wave Interference.
When some equipment may be the cause, turn off the power of
the equipment and check the functions of the MAGLUMI
®
Analyzer.If they are improved, Interference from the equipment is
probably the cause.
Try the following to correct.
a. Move the MAGLUMI
®
Analyzer further away from the
equipment that may be the cause.
b. Connect the power cord of the MAGLUMI
®
Analyzer to an
outlet that is on a different circuit from the equipment that
may be the cause.
c. Check that the other equipment, which is connected with this
Analyzer, is not affected by Electromagnetic Wave Interference.
Page 2