HRX-OM- T002
Chapter 3 Transport and Setting Up
3.3
Installation
HRS090 Series
3-20
3.3.5
Wiring of run/stop signal input
・
Remote signal input
Run/Stop signal input and remote signal input enable the product to
operate/stop or switched DIO REMOTE and DIO LOCAL remotely by
applying a contact signal input. This chapter illustrates examples of wiring.
Select DIO mode as the communication mode to activate the run/stop signal
input and remote signal input after wiring referring to Operation Manual
Communication Function.
[Tips]
This product has two input signals. These can be customized depending on
the customer’s application.
Table 3-6 Power supply, contact specifications
Name
Terminal NO.
Specification
Power supply output
5, 6, 7 (24VDC)
DC 24V ±10% 500mA MAX
1
13,14, 15 (24V COM)
Contact input signal 1
3 (Contact input signal 1)
-Run/stop
signal
input
-External
switch
signal input
2
Switch
the
input on the
operation
display panel.
Refer to the
Operation
manual
communication
function
for
details.
11 (Common of contact input signal 1)
Contact input signal 2
4 (Contact input signal 2)
-Run/stop
signal
input
-Remote
signal
input
-External
switch
signal input
2
12 (Common of contact input signal 2)
*1:
To use the power of the device, the total load current must be 500mA or less.
If the load is 500mA or more, the internal fuse will be cut to protect the product and the alarm [AL21 DC
line fuse cut] will be generated. Refer to Chapter 6 for handling of alarms.
*2: Refer to 3.3.6 Wiring of external switch signal input.
1.
Prepare the switch (power supply voltage: 24VDC, contact capacity: 35mA or more,
minimum load current: 5mA), and a signal cable (See