Final tests and measurements
149
851-164336 / D
8
Enter the measured hydrophone voltage
U(p-p)
into the
Measurement and calculation results
table, and calculate the
U Hydr
voltage in the same table.
→
Refer to table 6.
Measurements/calculations
Value
Unit
Example
Noise level
dB
--25.2
Hydrophone voltage
U(p--p)
volt
0.8
U(RMS) = U(p--p)/2
√
2
volt
0.28
U Hydr = 20 log U(RMS)
dB//1V
--10.9
Table 6 Measurement and calculation results
9
Make the requisite calculations in the
Measurement and
calculation results
table, and enter the result into the
Receiving voltage response (VR)
table.
→
The Receiving voltage response (VR) table is shown as
table 7.
10
Retrieve the
Hydrophone sensitivity as transmitter S
from
the
Test hydrophone data
table.
→
The Test hydrophone data table is shown as table 3 on page
144 (Source level measurements).
11
Enter the data into the
Receiving voltage response (VR)
table.
12
Retrieve the
20 log r
value from the
Measurements results
table.
→
The Measurements results table is shown as table 4 on page
146 (Source level measurements).
13
Enter the data into the
Receiving voltage response (VR)
table.
VR =
÷
(S + U Hydr
÷
20 log r)
Data
Value
Unit
Example
S = S Hydrophone
dB//
µ
Pa
/V
119.2
U Hydrophone
dB//1V
--10.9
20 log r
dB
17.5
VR (Voltage Response)
dB//1V/
µ
Pa
--90.8
Table 7 Receiving voltage response (VR)
14
Calculate the voltage response with the formula given in
Receiving voltage response (VR)
table.
Expected receiving voltage response is approximately:
•
VR =
-90
±
3 dB//1V/
µ
Pa
Summary of Contents for Simrad SP70
Page 2: ......
Page 14: ...Simrad SP70 X 851 164336 D This page is intentionally left blank...
Page 62: ...Simrad SP70 46 851 164336 D Figure 13 Mounting the Transceiver Unit CD3110A...
Page 226: ...Simrad SP70 210 851 164336 D Party Date Signature Party Date Signature Party Date Signature...
Page 227: ......
Page 229: ......