User Manual 63
:CALCulate:MARKer:PEAK:EXCursion?
Instructions
Sets/gets a relative threshold for the peak search criteria
Parameter Type
Float
,
unitdB
Parameter Range
0 ~ 200.0 dB
Return
Float
,
unitdB
Example
:CALCulate:MARKer:PEAK:EXCursion 10
Command format
:CALCulate:MARKer:PEAK:EXCursion:STATe
:CALCulate:MARKer:PEAK:EXCursion:STATe?
Instructions
Set and obtain the relative threshold switch
Parameter Type
Boolean
Parameter Range
OFF|ON|0|1
Return
0|1
Example
:CALCulate:MARKer:PEAK:EXCursion:STATe ON
3.7.12.3
Count Peak
Enable or disable continuous peak search. The default is Off. When enabled, the system will
always execute a peak search automatically after each sweep in order to track the signal under
measurement.
Command format
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe]
:CALCulate:MARKer[1]|2|3|4|5|6|7|8:CPSearch[:STATe]?
Instructions
Enable the continuous peak search function
Gets the status of the continuous peak search function switch
Parameter Type
Boolean
Parameter Range
OFF|ON|0|1
Return
0|1
Example
:CALCulate:MARKer1:CPSEarch ON
3.7.12.4
Peak Table
Open the peak table (in the lower window) which lists the peaks (with frequency and amplitude)
that meet the peak search condition. Up to 30 peaks can be displayed in the table.
Command format
:CALCulate:MARKer:PEAK:TABLe
:CALCulate:MARKer:PEAK:TABLe?