The measure average operation continuously calculates a weighted average for each index using
a specified number of averages.
Repeat
:
Each time the specified average is reached, the measurement resets the average counter.
Command format
[:SENSe]: TPOWer:AVERage:TCONtrol
[:SENSe]: TPOWer:AVERage:TCONtrol?
Instructions
Sets/gets average mode
Parameter Type
Enumeration
Parameter Range
EXPOnentialr|REPEat
Return
EXPOnentialr|REPEat
Example
:TPOWer:AVERage:TCONtrol REPE
3.9.6.2
Center Frequency
Set the center frequency, this center frequency which is the same with the center frequency of the
analyzer. Modifying this parameter will change the center frequency of the analyzer.
Command format
[:SENSe]:TPOWer:FREQuency:CENTer
[:SENSe]:TPOWer:FREQuency:CENTer?
Instructions
Set the time domain power center frequency
Obtain the time domain power center frequency
Parameter type
Float
,
unitHz
、
kHz
、
MHz
、
GHz
Parameter Range
related with Span
Return
Float
,
unitHz
Example
:TPOWer:FREQuency:CENTer 15kHz
3.9.6.3
Start line
Set the left margin (in time unit) of T-Power measurement. The data calculated under this
measurement is between the start line and stop line.
Command format
[:SENSe]:TPOWer:LLIMit
[:SENSe]:TPOWer:LLIMit?
Instructions
Set/Query the time domain power measurement start line
Parameter type
Float
,
unit : s