User Manual 151
Parameter type
Float
Parameter Range
10us~10ms
Return
Float
Example
:DDEMod:SYNC:BURSt:MINLength 0.5ms
5.1.4.5
Burst Min Gap
Setting the min gap for the rising and falling edges of the burst.
Represents the minimum distance (in "symbols") between adjacent bursts. The default value is 1
symbol in order to make sure that the burst search finds bursts that are very close to each other.
However, in case the "capture buffer" does not contain very close bursts, it is recommended that
you increase the value. This makes the burst search faster and also more robust for highly
distorted signals.
Note that this parameter only influences the robustness of the burst search. It should not be used
to explicitly exclude certain bursts from the measurement. For example, setting the minimum gap
length to 100 "symbols" does not ensure that the burst search does not find bursts that have a very
small gap.
Command format
[:SENSe]:DDEMod:SYNC:BURSt:MINGap
[:SENSe]:DDEMod:SYNC:BURSt:MINGap?
Instructions
Set the minimum burst interval
Gets the minimum burst interval
Parameter type
Float
Parameter Range
10us~10ms
Return
Float
Example
:DDEMod:SYNC:BURSt:MINGap 0.5ms
5.1.4.6
Sync Search
Synchronization search is to search for synchronization codewords in the measured signal, and
use this to segment and isolate the signal for subsequent display and analysis.
The synchronous codeword is a string of symbols (coding), so its length must be an integer
multiple of the number of bits (number of bits) of each symbol.
Command format
[:SENSe]:DDEMod:SYNC:SWORd[:STATe]
[:SENSe]:DDEMod:SYNC:SWORd[:STATe]?