188
User Manual
7.2.2
Preset Type
Perform the reset operation based on the current reset type. Select a configuration type for the
spectrum analyzer to reset and load.Reset Settings Preset types include default, last time, and
User.
⚫
Default: Reset the default loading parameters
⚫
Last time: reset to the state before the user's last software shutdown.
⚫
User: Resets and loads the configuration file specified by the user. This file can be obtained
by saving the user Settings.
Command format
:SYSTem:PRESet:TYPE DFT|LAST|USER
:SYSTem:PRESet:TYPE?
Note
Set the reset configuration type
Gets the reset configuration type
Parameter
type enumeration
Parameter Range
DFT
:
default
LAST
:
last
USER
:
user
The return value
enumeration
:
DFT|LAST|USER
The default value
DFT
Example
:SYSTem:PRESet:TYPE DFT
7.2.3
Save User Config
Saves the current system state as user-defined Settings to internal non-volatile storage.
Command format
:SYSTem:PRESet:USER#:SAVE
:SYSTem:PRESet:USER#:LOAD
LOAD
Save the user configuratio
Loading user Configuration
Example
:SYSTem:PRESet:USER2:SAVE
:SYSTem:PRESet:USER2:LOAD
Summary of Contents for SSA5083A