94
User Manual
3.9.7
TOI
Automatic measurement of IP3 (Third order Intercept Point), including the power of fundamental
wave and the Third order in the power, and calculate the adjustable Intercept Point.
Figure 3-12 TOI
TOI is an automatic measurement. There are no user controlled parameters.
Command format
:MEASure:TOI?
Instructions
Read the measurement results of third-order intermodulation distortion
Returns the following values separated by commas in scientific count form:
Lower Tone frequency (Hz), amplitude, Upper Tone frequency (Hz), amplitude,
TOI(Lower 3rd) frequency (Hz), amplitude, third-order intermodulation cutoff
(Intercept), High frequency TOI(Upper 3rd) frequency (Hz), amplitude, third-order
intermodulation cut-off point (Intercept).
Return
Float
Example
:MEASure:TOI?