7-1
7. MAINTENANCE
NOTE:
This section describes the maintenance procedures
that the service technician may perform.
In general the system does not require any
calibration.
7.1 Phantom Check
The service engineer should perform a routine check of the system
every time he visits the site, and when installing a system. In his
routine visit, the service engineer will check all the cables, connection
and general outlook of the system. The routine check will be performed
with the help of the service phantom.
(SEE “ACCESSORIES GUIDE” FOR INSTRUCTIONS ON THIS
CHECK)
Print out the phantom test at the end of the session and file it under
Register 7 of the system manual.
The phantom is supplied in a container filled with a solution. On the
side of the container is a level mark, indicating how much solution
must be in the container. If the solution does not reach the level mark,
solution must be added.
If the solution has become unclear, it must be replaced.
(See “Accessories guide”)
7.2 Monitor Check List
1. Click on the display icon from the control panel.
2. Select “SETTING” and click on “Test”.
3. Click on OK; the Test screen appears.
4. Click “Yes”/ OK.
5. Click on “Close” on the Control Panel.
7.3 Printer “Self Test”
To perform the printer “self test”, follow the instructions in the printer
manual under section test Menu.
SPB7-520.840.01.01.02
Summary of Contents for TS2000
Page 2: ...SPB7 520 840 01 01 02 ...
Page 4: ...SPB7 520 840 01 01 02 ...
Page 24: ...TS2000 Service Manual RD7409290 0 Rev F 2 14 SPB7 520 840 01 01 02 ...
Page 38: ...TS2000 Service Manual RD7409290 0 Rev F 4 10 SPB7 520 840 01 01 02 ...
Page 46: ...TS2000 Service Manual RD7409290 0 Rev F 5 8 SPB7 520 840 01 01 02 ...