6-1
6. TROUBLESHOOTING
NOTE:
1)
The effectiveness of any repair and part replacement must
be verified by a functional test. Refer to Section 7.4.
2)
Each intervention (replacement or repair) must be recorded
in the “record of service visits” section of the manual.
3) Replace “blown” fuses ONLY with the same type and
rating as marked on the safety label adjacent to the power
inlet, housing the fuse.
Be sure to unplug the power supply cord from the wall socket
BEFORE attempting to remove and/ or check the main power
fuse.
Every sophisticated system has the potential for an occasional problem to
occur. This section explains how to diagnose and correct these problems.
In the following table, “Reboot the system” means: use the exit button to shut
down the system properly, turn the power off, wait 4-5 seconds and then turn
the power on again.
6.1 Error Messages
6.1.1 Displayed Error Messages
Messages
Solution
Replay Parameters Not
Found.
•
Try replaying another patient file. If
still unable to complete the command,
reboot the system.
Probe Not Connected.
•
Check the cable connection for
probe(s). If the system still does not
respond appropriately, then reboot the
system.
•
Replace the probe.
Error 327: Has no...
•
Check that the parallel card is properly
installed.
Continued next page:
SPB7-520.840.01.01.02
Summary of Contents for TS2000
Page 2: ...SPB7 520 840 01 01 02 ...
Page 4: ...SPB7 520 840 01 01 02 ...
Page 24: ...TS2000 Service Manual RD7409290 0 Rev F 2 14 SPB7 520 840 01 01 02 ...
Page 38: ...TS2000 Service Manual RD7409290 0 Rev F 4 10 SPB7 520 840 01 01 02 ...
Page 46: ...TS2000 Service Manual RD7409290 0 Rev F 5 8 SPB7 520 840 01 01 02 ...