TS2000 Service Manual
RD7409290-0 Rev. F
6-8
3.
In the “Menu Bar," the “Tests” menu will appear.
4.
Click on the Stop/ Start button on the small probe/ large
probe; at the bottom of the screen, the message
“FPS=X.XX appears.
NOTE:
X.XX must be 8.00-+1.00 for the small probe.
Nominal value is 8.66
X.XX must be 2.00-+0.5 for the large probe.
Nominal value is 2.15
5.
Click again on the Start/ Stop button.
6.
In the “Tests” menu, click on “Back to User Menu”.
NOTE:
All other programs are for Research and
Development purposes only.
6.4.9 PM Board
On the PM, there are three red LEDs indicating presence of voltage. In
order to measure the voltages on the PM Board, measurements are
carried out between the ground of the output BNC and the following
points on the PM:
#
ID
LED
Value
1
L6
LED L1
+5V+/-5%
2
L4
LED L2
+12V+/-
5%
3
L3
LED L3
-12V+/-
5%
4
Zener Diode D9 (next to
U98)
-5V+/-5%
SPB7-520.840.01.01.02