TS2000 Service Manual
RD7409290-0 Rev. F
6-6
6.4 PM and PSB TESTS
6.4.1 Preparation
1.
Exit system to Windows.
2.
Access the Windows NT Explorer.
3.
Double-click
PM Test
in the TransScan directory.
6.4.2 Serial Communication Test
1.
Perform Preparation 6.4.1.
2.
Click
PC2PM
to check serial communication between
PM and computer.
3.
Click
HELLO
in the menu.
4.
On the screen, two response messages will appear (in
bold letters). If this does not occur, refer to
“Communication Error” in section 6.1.
5.
Select “View” and click
Clear
.
6.
Select “Single” and click
Probe
.
7.
Check that:
A
is selected in the “Interf” box/ small is
selected in the “Probe” box.
8.
Click
Read
to get the ID of the PSB.
9.
PSB ID =’03’ will appear in binary form ‘0011’ in the
lower Data box. (Two checks in the last digits)
10.
Click
Cancel
and repeat Step 5.
6.4.3 Check for the P.S.B: Socket (Port I.D.)
1.
Insert the address 001 in the b6-b4 box in the lower
control section and connect a probe to socket J1 in the
PSB.
2.
Click on the “Read” button; the address 0001 appears in
the data section (b7-b0).
3.
Move the probe to socket J2 and click on the “Read”
button; the address will be 0010.
4.
Move the probe to socket J3 and click on the “Read”
button; the address will be 0100.
5.
Clear the b6-b4 box. Do not clear to read.
SPB7-520.840.01.01.02
Summary of Contents for TS2000
Page 2: ...SPB7 520 840 01 01 02 ...
Page 4: ...SPB7 520 840 01 01 02 ...
Page 24: ...TS2000 Service Manual RD7409290 0 Rev F 2 14 SPB7 520 840 01 01 02 ...
Page 38: ...TS2000 Service Manual RD7409290 0 Rev F 4 10 SPB7 520 840 01 01 02 ...
Page 46: ...TS2000 Service Manual RD7409290 0 Rev F 5 8 SPB7 520 840 01 01 02 ...