TS2000 Service Manual
RD7409290-0 Rev. F
6-2
Messages
Solution
Parallel card has no number
•
Make sure the program was activated.
Modem Error
•
Check power to the modem
A message on system startup
•
Check the connection between the
modem and the computer.
•
Make sure the modem is set to start up
with the system
Error Message: PM is not
responding to message 6...
•
Check the Serial Communication cable
connection in the back of the system.
•
Check that the 3 LED’s on the P.M.
board are ON
•
Check the voltages on the J2 connector
on the PM. The voltages should be 5
Volts, +12 and -12 Volts.
•
Check that the setup for Com1 is
115200.
•
Check parallel cable connected the PM
adapter to the I/O card.
Amplitude Error / over
current
•
Check that there are no foreign metals
on the probe.
•
Check the connection cable between
the PM and the handheld cylinder.
•
Press the "GO" button on the probe
again. If the system is still not
responding appropriately, reboot the
system.
6.1.2 TransScan TS 2000 Error Message
1.
Enter “Database."
2.
Select “Service”/ “Errors” and Click “Print Last 10
Logs."
SPB7-520.840.01.01.02