ID
Icons
Message
Cause/Action
206
CH3 pulse overflow
Pulse output has insufficient pulse separation. Increase
Amount per pulse, or reduce Pulse width, or reduce Pulses
per amount.
207
CH4 loop current in lower
saturation
The process value is so low that the loop current reaches
the lower saturation limit and cannot decrease any further.
Check process conditions. Adjust parameter value 'Lower
range value'.
208
CH4 loop current in upper
saturation
The process value is so high that the loop current reaches
the upper saturation limit and cannot increase any further.
Check process conditions. Adjust parameter value 'Upper
range value'.
209
CH4 cable break
Check current output cable connection. Passive operation:
check external power supply.
210
CH4 output frequency too
low
The process value is below the parameter 'Lower range
value'. The output frequency cannot decrease any further.
Check process conditions. Adjust parameter value 'Lower
range value'.
211
CH4 output frequency too
high
The process value is above the parameter 'Upper range
value'. The output frequency cannot increase any further.
Check process conditions. Adjust parameter value 'Upper
range value'.
212
CH3 pulse overflow
Pulse output has insufficient pulse separation. Increase
Amount per pulse, or reduce Pulse width, or reduce Pulses
per amount.
213
Invalid dosing settings
Correct settings.
214
CH2 simulated
Disable the simulation to return to normal operation.
Diagnostics and troubleshooting
10.2 Fault codes and corrective actions
SITRANS FC330
Operating Instructions, 01/2019, A5E44030648-AB
187